Elekta, Inc. began a recall of Leksell SurgiPlan on . The reason FDA records is "Images from a different patient can be imported without a warning for the user". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1428-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Leksell SurgiPlan
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Images from a different patient can be imported without a warning for the user.
- FDA root cause
- Other
- Action
- Phase I - A technical note for LGP and SPS describing the risk for importing wrong image files and including recommendations on how to reduce this risk has been written and distributed to users of both Leksell GammaPlan and Leksell SurgiPlan on May 23, 2005 via FCO DOC 05-0001. Phase 2 - A maintenance release was created of the Scanner Interface (SIF) to correct the problem. This was distributed to Leksell SurgiPlan users on November 30, 2005 via FCO0000153-1. Additional questions are directed to the firm to Product Manager lrobert.aslund AT elekta.com, or After Sales gustaf.piehl AT elekta.com..
- Quantity in commerce
- 57 units
- Distribution
- Nationwide, Canada and Mexico.
- Product codes and lots
- US68104248, SG50425261, US68154313, US66320311, 3415G15729, 3416G05185, SW ONLY, 3416G07810, 3416G08464, US19311023, SG5121735, SG44720158, 000000000000010, SG33221023, 000000000000015, SG41020267, US19477688, SG14720637, SG31020878, SG21220437, IE24488098, SG11720734, US6836799, SG20220679, US68064889, US19344670, SG41720337, SG44820136, US6863578, US68247174, SG20220701, 000000000000011, 6458A30012, HP715/80, 000000000000019, 3628G06513, SG5020708, SG31120383, LGP-SG13420315, 000000000000014, 00000000000004, 3711C04231, US68154724, C160, SG33820630, 00000000000002, 000000000000013, 6146A06124, SG41920535, 6513A30173, 0304011939, A4986A, SG20220694, 000000000000016, SG20220699, 6601A30228, US19476876