EasyRA Cuvette Segment 1, P/N 007000-001 (In Vitro Diagnostic ) EasyRA Cuvette Segments included in the...
Medica Corporation began a recall of EasyRA Cuvette Segment 1, P/N 007000-001 (In Vitro Diagnostic ) EasyRA Cuvette Segments included in the following Kits: EasyRA Cuvettes, REF 10700-4 Startup... on . The reason FDA records is "Chemistry analyzers may report Calcium results below the sample's actual concentration. Reported results may be ten to fifteen percent lower than expected sample results. The defect is a shortened path length at...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0078-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medica Corp.
- Product
- EasyRA Cuvette Segment 1, P/N 007000-001 (In Vitro Diagnostic ) EasyRA Cuvette Segments included in the following Kits: EasyRA Cuvettes, REF 10700-4 Startup Kit - Chemistry, REF 10990 Startup Kit - Chemistry & ISE, REF 10991
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Chemistry analyzers may report Calcium results below the sample's actual concentration. Reported results may be ten to fifteen percent lower than expected sample results. The defect is a shortened path length at specific cuvette positions of those segments manufactured using molding cavity 1. The shortened path length was determined by measuring the cuvette dimensions with a Co-ordinate Measuri
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Medica issued an Urgent Product Recall letter dated March 29, 2010 to customers. Users were requested to cease using and discard the affected product, and complete and return a Response Form in order to obtain replacement product. Customers can contact Medica Technical Support at 877-777-5895.
- Quantity in commerce
- 14,926
- Distribution
- Worldwide Distribution: USA, and the countries of Bulgaria, Canada, India, Mexico, Panama, Phillipines, Russia, Turkey, New Zealand, South Africa, United Arab Emirates, and Venezuela.
- Product codes and lots
- Lot Numbers: 2009/A2810, 2009/A2910, 2009/A3010, 2010/A2001, 2010/A21 01, 2010/A1901, 2009/A0211, 2009/A0311