The magnetic headrest is designed to fit the following dental chairs (THE HEADREST IS THE RECALLED ITEM, NOT...
Dental Equipment LLC began a recall of The magnetic headrest is designed to fit the following dental chairs (THE HEADREST IS THE RECALLED ITEM, NOT THE CHAIRS): The dental chairs are sold under the... on . The reason FDA records is "The magnetic headrest on certain dental chairs may affect the function/programming of some implantable pacemakers or defibrillators if the implanted device is programmed to respond to a magnet". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2241-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DCI International, LLC
- Product
- The magnetic headrest is designed to fit the following dental chairs (THE HEADREST IS THE RECALLED ITEM, NOT THE CHAIRS): The dental chairs are sold under the brand names Marus and DCI Equipment. The affected models are: Marus MaxStar-Agency Model # DC1490; Marus MaxStar-Agency Model # DC1690; Marus NuStar-Agency Model # DC1700; Marus NuStar-Agency Model # DC1702; Marus ProStar-Agency Model # DC1540; Marus ProStar-Agency Model # DC1535; DCI Equipment DC1235 Hydraulic-Agency Model # DC1235; and DCI Equipment DC1335 Electromechanical-Agency Model # DC1335.
- Product code
- NRU Chair, Dental, Without Operative Unit
- Reason for recall
- The magnetic headrest on certain dental chairs may affect the function/programming of some implantable pacemakers or defibrillators if the implanted device is programmed to respond to a magnet.
- FDA root cause
- Other
- Action
- DCI Equipment sent an "Urgent Field Safety Alert" letter dated March 15, 2010 to consignees advising them of the potential problem. The letter instructed that patients with implantable pacemakers or defibrillators should avoid dental chairs with the magnetic headrest. A warning label was sent with the letter for insertion into the product Use and Care manual for operator warning. Consignees were asked to complete a Field Safety Alert Acknowledgement/Return Form and fax to the firm. Consignees were requested to notify customers.Consignees may contact the firm at 503 537-3602.
- Quantity in commerce
- 108
- Distribution
- Worldwide distribution in the USA and Canada.
- Product codes and lots
- The individual headrests do not have serial numbers. The headrest has a part number of "40R501". The part number is followed by a dash and a 2 or 3 digit number that indicates the color of the vinyl covering.