Z-0071-2011

Continuum Acetabular Systems Shell Inserter Adapter with Rotational Control, 00-8790-003-00, Zimmer Inc...

Zimmer GmbH / initiated 2010-03-19 / Terminated / root cause: design

Zimmer Inc. began a recall of Continuum Acetabular Systems Shell Inserter Adapter with Rotational Control, 00-8790-003-00, Zimmer Inc., Warsaw, IN. Used to mate the inserter with the shell... on . The reason FDA records is "The firm has received reports of users encountering difficulties in usage leading to surgical delays or minor damage to a threaded surface of the shell implant". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0071-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Continuum Acetabular Systems Shell Inserter Adapter with Rotational Control, 00-8790-003-00, Zimmer Inc., Warsaw, IN. Used to mate the inserter with the shell implant.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The firm has received reports of users encountering difficulties in usage leading to surgical delays or minor damage to a threaded surface of the shell implant.
Root cause tag
design
FDA root cause
Device Design
Action
The firm sent Urgent: Device Removal letters dated 3/15/2010. The letter identified the affected product, described the issue, discussed the clinical implications, and provided a list of required actions. The consignees were instructed to perform an inventory of the recalled product and report the amount on hand to Zimmer by completing the Inventory Return Certification Form. After Zimmer received the report, they would send out new product to the consignee. After new product was received the consignees were instructed to return the recalled product to Zimmer. The firm expects to have all product replaced by 4/30/2010. If there are any questions or additional info needed, consignees should contact Zimmer, Inc. at 1-800-613-6131 or 574-372-4463.
Quantity in commerce
1125 all products
Distribution
Worldwide Distribution -- US, Singapore, Australia, Portugal, Germany, Switzerland, Spain, Finland, United Kingdom, Israel, Italy and Sweden.
Product codes and lots
61301130, 61309997, 61326687, 61330544, 61389660, 61407677 and 61426853.