Z-1537-2010
Medtronic HMS PLUS, Hemostasis Management System. Medtronic, Inc., Cardiovascular Division, Moundsview, MN...
Medtronic, Inc. / initiated 2010-03-08 / Terminated
Medtronic Inc. Cardiac Rhythm Disease Managment began a recall of Medtronic HMS PLUS, Hemostasis Management System. Medtronic, Inc., Cardiovascular Division, Moundsview, MN 55112. A microprocessor based, multichannel clot... on . The reason FDA records is "The heparin assay controls may run longer than normal and in some cases, fail to give an acceptable result". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1537-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic HMS PLUS, Hemostasis Management System. Medtronic, Inc., Cardiovascular Division, Moundsview, MN 55112. A microprocessor based, multichannel clot timing instrument with automated syringe handling for pipetting blood into single use cartridges. It performs in vitro heparin sensitivity evaluations, heparin assays, activated clotting times and platelet function evaluations.
- Product code
- JPA System, Multipurpose For In Vitro Coagulation Studies
- Reason for recall
- The heparin assay controls may run longer than normal and in some cases, fail to give an acceptable result.
- FDA root cause
- Process change control
- Action
- Medtronic issued an "Urgent Medical Device Notification" letter dated March 8, 2010 addressed to "Dear Valued Customer" and was sent to three locations at each customer site: Risk Management, Cardiovascular Surgery and Clinical Laboratory-Point of Care. The letter described the product and the issue, provided instructions for control and patient testing. The customer was requested to return an enclosed certificate as proof of notification. For further information, contact your Medtronic Sales Representative or call 1-763-391-9915.
- Quantity in commerce
- 508 US, 36 OUS
- Distribution
- Worldwide Distribution -- United States, International West (countries included in Western Europe and the Middle East), Australia and Canada.
- Product codes and lots
- HMS Plus Catalog Number: 30514; Heparin Assay Controls: 306-01POR, 306-02POR, 306-03POR, 306-04POR, 306-05POR and 306-09POR.