Z-1433-2010
3DKNEE SYSTEM, 3D Knee Tibial Insert-DCM, Sterile, Right, size 4 x 11mm. Encore Medical, LP. Austin, TX...
Encore Medical, LP / initiated 2010-03-19 / Terminated / root cause: packaging
Encore Medical, Lp began a recall of 3DKNEE SYSTEM, 3D Knee Tibial Insert-DCM, Sterile, Right, size 4 x 11mm. Encore Medical, LP. Austin, TX 78758-5445. Intended to aid the surgion in relieving... on . The reason FDA records is "Affected product had incorrect color label which is a secondary visual indicator for sizing". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1433-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- 3DKNEE SYSTEM, 3D Knee Tibial Insert-DCM, Sterile, Right, size 4 x 11mm. Encore Medical, LP. Austin, TX 78758-5445. Intended to aid the surgion in relieving the patient of knee pain and restoring knee joint function.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Affected product had incorrect color label which is a secondary visual indicator for sizing.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Encore Medical, LP notified consignees by e-mail dated March 19, 2010. Consignees were requested to return the affected product to the firm. For further information, contact Encore Medical, LP at 1-888-838-3668.
- Quantity in commerce
- 7 units
- Distribution
- Worldwide Distribution -- United States (UT, IA) and Germany.
- Product codes and lots
- Catalog Number: 392-11-604, Lot Number: 59600085.