Unicel DxH 800 Coulter Cellular Analysis System. Beckman Coulter, Inc. Brea, CA. Quantitative, automated...
Beckman Coulter Inc. began a recall of Unicel DxH 800 Coulter Cellular Analysis System. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer and for in vitro diagnostic use in... on . The reason FDA records is "Issue 1: The system will substitute or omit the characters # @ 1 \ ] ' ( I ) - when the instrument is configured in a language other than English, Chinese or Japanese. Example: Using the character @, a substitution in...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1658-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Unicel DxH 800 Coulter Cellular Analysis System. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer and for in vitro diagnostic use in screening patient populations found in clinical laboratories.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- Issue 1: The system will substitute or omit the characters # @ 1 \ ] ' ( I ) - when the instrument is configured in a language other than English, Chinese or Japanese. Example: Using the character @, a substitution in the string '123@456' would result in '123#456' and an omission would result in '123456'. Issue2: The characters 7 " and * (question mark, quotation mark, and asterisk) used as part
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Beckman Coulter, Inc. issued "Urgent: Product Corrective Action" notifications dated February 12, 2010. Consignees were informed of the affected product and instructed on proper actions. The firm also asked that the enclosed response forms be returned. For further information, contact your local Beckman Coulter Representative or call 1-800-526-7694.
- Quantity in commerce
- 56 units
- Distribution
- Distributed Nationwide and Canada.
- Product codes and lots
- Part number: 629029; All Serial Numbers.