Z-1658-2010

Unicel DxH 800 Coulter Cellular Analysis System. Beckman Coulter, Inc. Brea, CA. Quantitative, automated...

Siemens Healthcare Diagnostics Inc. / initiated 2010-02-04 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of Unicel DxH 800 Coulter Cellular Analysis System. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer and for in vitro diagnostic use in... on . The reason FDA records is "Issue 1: The system will substitute or omit the characters # @ 1 \ ] ' ( I ) - when the instrument is configured in a language other than English, Chinese or Japanese. Example: Using the character @, a substitution in...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1658-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Unicel DxH 800 Coulter Cellular Analysis System. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer and for in vitro diagnostic use in screening patient populations found in clinical laboratories.
Product code
GKZ Counter, Differential Cell
Reason for recall
Issue 1: The system will substitute or omit the characters # @ 1 \ ] ' ( I ) - when the instrument is configured in a language other than English, Chinese or Japanese. Example: Using the character @, a substitution in the string '123@456' would result in '123#456' and an omission would result in '123456'. Issue2: The characters 7 " and * (question mark, quotation mark, and asterisk) used as part
Root cause tag
software
FDA root cause
Software design
Action
Beckman Coulter, Inc. issued "Urgent: Product Corrective Action" notifications dated February 12, 2010. Consignees were informed of the affected product and instructed on proper actions. The firm also asked that the enclosed response forms be returned. For further information, contact your local Beckman Coulter Representative or call 1-800-526-7694.
Quantity in commerce
56 units
Distribution
Distributed Nationwide and Canada.
Product codes and lots
Part number: 629029; All Serial Numbers.