Z-0456-2011

i-STAT CREA Cartridge, Abbott List Number: 06F10-01, 02; IVD. Manufactured by Abbott Point of Care, Inc...

Abbott Medical / initiated 2010-03-17 / Terminated / root cause: labeling and UDI

Abbott Point of Care Inc. began a recall of i-STAT CREA Cartridge, Abbott List Number: 06F10-01, 02; IVD. Manufactured by Abbott Point of Care, Inc., Princeton, NJ on . The reason FDA records is "The i-STAT CREA Cartridges may produce inaccurate (depressed) results at or near the end of the labeled expiry date when stored above 25 degrees for 2 weeks". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0456-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
i-STAT CREA Cartridge, Abbott List Number: 06F10-01, 02; IVD. Manufactured by Abbott Point of Care, Inc., Princeton, NJ.
Product code
CGL Electrode, Ion Based, Enzymatic, Creatinine
Reason for recall
The i-STAT CREA Cartridges may produce inaccurate (depressed) results at or near the end of the labeled expiry date when stored above 25 degrees for 2 weeks.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Abbott Point of Care issued Urgent Product Recall Notification letters dated March 2010 to all direct customers identifying the affected product and the actions to be taken. Customers were instructed to determine if they had cartirdges that had been stored above 25 degrees Celsius, and if so, contact Abbott for replacement or credit. Abbott Technical Support can be contacted at 800-366-8020 option 1 about this recall.
Quantity in commerce
34525 cartridges
Distribution
Worldwide Distribution: Throughout USA and Canada.
Product codes and lots
Lot numbers A09279A and A09280A, Exp March 2010