Z-1434-2010

Orthofix US XCaliber Meta-Diaphyseal Complete Kit, Sterile. 99-91600, manufactured by Orthofix SRL, Via delle...

Orthofix SRL / initiated 2010-02-22 / Terminated / root cause: manufacturing process

Orthofix Srl began a recall of Orthofix US XCaliber Meta-Diaphyseal Complete Kit, Sterile. 99-91600, manufactured by Orthofix SRL, Via delle Nazioni, 9, 27012 Bussolengo (VR) Italy. Intended... on . The reason FDA records is "During standard quality controls, firm identified a potential for breakage of the XCaliber clamp when locking the cam". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1434-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthofix SRL
Product
Orthofix US XCaliber Meta-Diaphyseal Complete Kit, Sterile. 99-91600, manufactured by Orthofix SRL, Via delle Nazioni, 9, 27012 Bussolengo (VR) Italy. Intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation and other bone conditions amenable to treatment by use of the external fixation modality.
Product code
JDW Pin, Fixation, Threaded
Reason for recall
During standard quality controls, firm identified a potential for breakage of the XCaliber clamp when locking the cam.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Orthofix issued an "Urgent: Field Safety Notice" dated February 22, 2010. Consignees were informed of the affected product and instructed to cease use and to return the product to the firm.
Quantity in commerce
9
Distribution
Nationwide Distribution -- TX, ND, NY, FL and CA.
Product codes and lots
Lot Numbers: IT965408