Z-1435-2010
Orthofix US XCaliber Ankle Complete Kit, Sterile, 99-91647. manufactured by Orthofix SRL, Via delle Nazioni...
Orthofix SRL / initiated 2010-02-22 / Terminated / root cause: manufacturing process
Orthofix Srl began a recall of Orthofix US XCaliber Ankle Complete Kit, Sterile, 99-91647. manufactured by Orthofix SRL, Via delle Nazioni, 9, 27012 Bussolengo (VR) Italy. Intended for use... on . The reason FDA records is "During standard quality controls, firm identified a potential for breakage of the XCaliber clamp when locking the cam". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1435-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Orthofix SRL
- Product
- Orthofix US XCaliber Ankle Complete Kit, Sterile, 99-91647. manufactured by Orthofix SRL, Via delle Nazioni, 9, 27012 Bussolengo (VR) Italy. Intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation and other bone conditions amenable to treatment by use of the external fixation modality.
- Product code
- JDW Pin, Fixation, Threaded
- Reason for recall
- During standard quality controls, firm identified a potential for breakage of the XCaliber clamp when locking the cam.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Orthofix issued an "Urgent: Field Safety Notice" dated February 22, 2010. Consignees were informed of the affected product and instructed to cease use and to return the product to the firm.
- Quantity in commerce
- 25
- Distribution
- Nationwide Distribution -- TX, ND, NY, FL and CA.
- Product codes and lots
- Lot Numbers: IT956155 and IT967331