Z-1602-2010
Polarus(R) Humeral Fixation System, size: 3.5mm x 25.0mm Cortical Screw, QTY 1, Ref: HCO3250-S Acumed, LLC...
Acumed LLC / initiated 2009-05-19 / Terminated / root cause: manufacturing process
Acumed LLC began a recall of Polarus(R) Humeral Fixation System, size: 3.5mm x 25.0mm Cortical Screw, QTY 1, Ref: HCO3250-S Acumed, LLC Hillsboro, OR Intended use: Acumed intramedullary... on . The reason FDA records is "Mix-up of parts - Product labeled cortical bone screw actually contains Acutract fixation screw". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1602-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Acumed LLC
- Product
- Polarus(R) Humeral Fixation System, size: 3.5mm x 25.0mm Cortical Screw, QTY 1, Ref: HCO3250-S Acumed, LLC Hillsboro, OR Intended use: Acumed intramedullary rods and screws are designed to provide fixation of humeral, forearm and fibula fractures while they heal.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- Mix-up of parts - Product labeled cortical bone screw actually contains Acutract fixation screw
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- On May 18, 2009, the firm, ACUMED, began calling customers. On May 19, 2009, the ACUMED sent e-mails and a "URGENT NOTICE: DEVICE RECALL" Initial Notice letter to all customers. The letter describes the product, problem and actions to be taken by customers. The customers were instructed to identify the product having the lot code(s) and quarantine them, return all products from the lot code(s) to Acumed immediately, to please notify the customers of this recall if you further distributed this product, and complete and return the Acumed Product Recall Effectiveness Form. Please contact Acumed Customer Service at 1-888-627-9957 for any questions regarding this notification.
- Quantity in commerce
- 83
- Distribution
- Worldwide distribution: USA including FL, GA, MI, MN, NY, TX, VA and countries of United Kingdom, Herzeles, South Korea, South Africa, China, and Milan
- Product codes and lots
- Lot code 203214