Z-1944-2010
Tabs Professional Monitor, includes one wall bracket, Model 25022. Stanley Senior Technologies, Lincoln, NE
Securitas Healthcare LLC / initiated 2010-03-26 / Terminated
Stanley Security Solutions, Inc. began a recall of Tabs Professional Monitor, includes one wall bracket, Model 25022. Stanley Senior Technologies, Lincoln, NE on . The reason FDA records is "Product might not indicate low battery alarms or exit alarms when powered by a DC power supply". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1944-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Securitas Healthcare LLC
- Product
- Tabs Professional Monitor, includes one wall bracket, Model 25022. Stanley Senior Technologies, Lincoln, NE.
- Product code
- KMI Monitor, Bed Patient
- Reason for recall
- Product might not indicate low battery alarms or exit alarms when powered by a DC power supply.
- FDA root cause
- Component change control
- Action
- The firm made phone calls beginning 3/26/10 to customers who purchased the 12-volt power supply informing them of the recall and instructing them to determine the number of units that will need to be returned for replacement. Important Safety Notice letters dated 3/25/10 were issued via regular mail on 3/26/10 and also providing the reason for the recall, instructions, and customer service contact information to arrange for return of the units.
- Quantity in commerce
- 84 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Serial numbers 235388 thru 237658.