Z-1945-2010

Tabs Professional Monitor Without Wall Bracket, Model 25023. Stanley Senior Technologies, Lincoln, NE

Securitas Healthcare LLC / initiated 2010-03-26 / Terminated

Stanley Security Solutions, Inc. began a recall of Tabs Professional Monitor Without Wall Bracket, Model 25023. Stanley Senior Technologies, Lincoln, NE on . The reason FDA records is "Product might not indicate low battery alarms or exit alarms when powered by a DC power supply". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1945-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Securitas Healthcare LLC
Product
Tabs Professional Monitor Without Wall Bracket, Model 25023. Stanley Senior Technologies, Lincoln, NE.
Product code
KMI Monitor, Bed Patient
Reason for recall
Product might not indicate low battery alarms or exit alarms when powered by a DC power supply.
FDA root cause
Component change control
Action
The firm made phone calls beginning 3/26/10 to customers who purchased the 12-volt power supply informing them of the recall and instructing them to determine the number of units that will need to be returned for replacement. Important Safety Notice letters dated 3/25/10 were issued via regular mail on 3/26/10 and also providing the reason for the recall, instructions, and customer service contact information to arrange for return of the units.
Quantity in commerce
36 units
Distribution
Nationwide Distribution
Product codes and lots
Serial numbers 236662 thru 236697.