Freestyle Navigator Continuous Glucose Monitor System, ART 16467 Rev A, System Kit MMOL EMEA. Abbott Item...
Abbott Diabetes Care, Inc. began a recall of Freestyle Navigator Continuous Glucose Monitor System, ART 16467 Rev A, System Kit MMOL EMEA. Abbott Item Number 71036-01, distributed by Abbott Diabetes Care... on . The reason FDA records is "Potential failure or inaccurate readings--the device's plastic housing near the battery compartment may crack, and allow moisture to enter". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1455-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Freestyle Navigator Continuous Glucose Monitor System, ART 16467 Rev A, System Kit MMOL EMEA. Abbott Item Number 71036-01, distributed by Abbott Diabetes Care Inc, Alameda, CA. Medical device indicated for continually recording interstitial fluid glucose levels in people ages 18+ with diabetes mellitus for improving diabetes management. The device is indicated to be used in hospital and home settings to aid health care.
- Product code
- MDS Sensor, Glucose, Invasive
- Reason for recall
- Potential failure or inaccurate readings--the device's plastic housing near the battery compartment may crack, and allow moisture to enter.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Notification letters were issued to FreeStyle Navigator customers via certified mail and notification was posted on the Abbott Diabetes Care's Website. Customers were notified of the potential issue and actions to take: 1) if moisture is observed in the FreeStyle Navigator transmitter now or at any time, discontinue use of the transmitter and call Abbott Diabetes Care Customer Care at 1-800-418-6427 and 2) if a loss of connection is encountered or if your results are not reliable or are inconsistent with how you feel, consult your User's Guide for troubleshooting guidelines. If the issue is not resolved, call Abbott Diabetes Care Customer Care at 1-800-418-6421. For outside-of-US accounts, the firm's regional offices will distribute notices to identified customers in the appropriate language.
- Quantity in commerce
- 5,449 of total units
- Distribution
- Worldwide Distribution -- United States (OH, FL, NJ, CA, MS, MN, MA, and NY), France, Australia, the Netherlands, Sweden, the UK, Israel, Ireland and Germany.
- Product codes and lots
- Abbott Item # 71036-01 with multiple lot numbers, kits serial numbers, and Tx serial numbers. Lot numbers are these: 01N0158 or 01N0278 or 01N0348 or 01N0568 or 02N0128. Kit Serial numbers beginning with: BAAH 156 BAAH 162 or BAAH 200 or BAAH 207 or BAAH 220 or BAAH 223 or BAAH 275 or BAAH 290 and followed by a five (5 )digit number. Tx Serial numbers beginning with: AAAH 207 or AAAH 212 or AAAH 221 or AAAH 228 or AAAH 268 or AAAH 283 or AAAH 296, and AAAK 008 and followed by a five (5) digit number.