ConMed Goldline Hand Controlled With Blade Electrode, Tip Cleaner, Rocker Switch, Safety Holster and 10 FT...

ConMed Corporation / initiated 2010-03-29 / Terminated / root cause: design

ConMed Electrosurgery began a recall of ConMed Goldline Hand Controlled With Blade Electrode, Tip Cleaner, Rocker Switch, Safety Holster and 10 FT Cable, REF 130304A, Sterile R, Assembled in Mexico... on . The reason FDA records is "Electrosurgical pencil may remain activated or self activate after power switch is released". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1563-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConMed Corporation
Product
ConMed Goldline Hand Controlled With Blade Electrode, Tip Cleaner, Rocker Switch, Safety Holster and 10 FT Cable, REF 130304A, Sterile R, Assembled in Mexico, ConMed Corporation, 525 French Rd., Utica, NY 13502.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Electrosurgical pencil may remain activated or self activate after power switch is released.
Root cause tag
design
FDA root cause
Device Design
Action
Customers were notified by letter on March 25, 2010, and were instructed to return all affected devices and to forward the recall information to anyone that may have received the devices to the end user. More information is available by contacting the firm directly at 800-552-0138.
Quantity in commerce
13,920 units (US), 2,080 units (Foreign)
Distribution
Nationwide, including 2 VA facilities in GA and 1 military facility in UT. Foreign distribution to Australia, Belgium, Canada, China, Dominican Republic, El Salvador, Japan, Korea, Mexico, Puerto Rico, Saudi Arabia, Spain, Taiwan.
Product codes and lots
Lot code range: 0903253 through 1002143