Z-1524-2010
Ascension PyroCarbon PIP Total Joint - Trials; Ascension PIP, sizes 10P, 20P, 30P and 40P. Ascension...
Ascension Orthopedics, Inc. / initiated 2009-06-17 / Terminated / root cause: design
Ascension Orthopedics, Inc began a recall of Ascension PyroCarbon PIP Total Joint - Trials; Ascension PIP, sizes 10P, 20P, 30P and 40P. Ascension Orthopedics, 8700 Cameron Road, Suite 100, Austin, Texas... on . The reason FDA records is "Trial product was larger than actual implant. Trial product was re-designed to more closely match the implant". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1524-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ascension Orthopedics, Inc.
- Product
- Ascension PyroCarbon PIP Total Joint - Trials; Ascension PIP, sizes 10P, 20P, 30P and 40P. Ascension Orthopedics, 8700 Cameron Road, Suite 100, Austin, Texas 78754; TEL: 512 836-5001. Intended as an accessory to the Ascension PyroCarbon PIP Total Joint used in finger joint implant procedures where the joint has become painful, stiff or cannot move because of arthritis.
- Product code
- NEG Finger Semi-Constrained Pyrolytic Carbon Uncemented Prosthesis
- Reason for recall
- Trial product was larger than actual implant. Trial product was re-designed to more closely match the implant.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Ascension Orthopedics, Inc. issued e-mail notification to distributors and sales representatives beginning June 2009. Consignees were instructed return affected product to the firm for replacement. For further information, contact Ascension Orthopedics, Inc. at 1-512-836-5001 extension 1549.
- Quantity in commerce
- 129 units (64 US, 65 OUS)
- Distribution
- Worldwide Distribution - Worldwide Distribution - United States, United Kingdom, France, Australia, Belgium, Czech Republic, Canada, Germany, The Netherlands and South Africa.
- Product codes and lots
- All lot codes.