Z-2658-2010

Access Immunoassay Systems Free T4, Part Number: 33880

Siemens Healthcare Diagnostics Inc. / initiated 2010-03-24 / Terminated / root cause: labeling and UDI

Beckman Coulter Inc. began a recall of Access Immunoassay Systems Free T4, Part Number: 33880 on . The reason FDA records is "The recall was initiated after Beckman Coulter determined that kit boxes for Access Free T4 reagent (REF 33880), lot number 916293 may contain reagent packs labeled as Access Vitamin B12, (REF 33000), lot 916277. Using...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2658-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Immunoassay Systems Free T4, Part Number: 33880
Product code
CEC Radioimmunoassay, Free Thyroxine
Reason for recall
The recall was initiated after Beckman Coulter determined that kit boxes for Access Free T4 reagent (REF 33880), lot number 916293 may contain reagent packs labeled as Access Vitamin B12, (REF 33000), lot 916277. Using an Access Vitamin B12 reagent pack, lot 916277, filled with Access Free T4 reagents may generate incorrect patient results.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
A Product Corrective Action (PCA) letter was sent on March 24, 2010 to the affected customers. The letter provides the customers with an explanation of the problem identified and instructs them to discard the FT4 kit, Lot 916293 in their stock if the FT4 packs within the kit packaging are labeled as Vitamin B12 Lot 916277. The PCA requested customers: (1) Inspect all Access Vitamin B12 reagent packs that have been removed from their kit boxes, including reagent packs currently loaded on your system and discard any reagent packs labeled as Access Vitamin B12, lot 916277; (2) Inspect your Access Free T4 reagent kit inventory for any reagent packs labeled as Access Vitamin B12, lot 916277 and discard any reagent packs labeled as Access Vitamin B12, lot 916277; (3) Review your historical quality control to ensure that assay performance for Access Vitamin B12 lot 916277 has been within your laboratory specifications. In addition customers were instructed to share this information with their laboratory staff and retain this notification as part of your laboratory Quality System documentation. Complete and return the enclosed response form within 10 days so Beckman Coulter can be assured that you have received this important notification. If customers need assistance or have any questions regarding the notification, they were instructed to contact Customer Technical Support Center at 1-800-854-3633 in the United States and Canada.
Quantity in commerce
10826
Distribution
Nationwide In the US & Canada
Product codes and lots
Lot Number 916293