Z-1536-2010

Bard Ventrio Small Oval Hernia Patch 8cm X 12cm (3.1" x 4.7"), Warwick, RI

Davol Inc., Sub. C.R. Bard, Inc. / initiated 2010-04-08 / Terminated / root cause: labeling and UDI

Davol, Inc., Sub. C. R. Bard, Inc. began a recall of Bard Ventrio Small Oval Hernia Patch 8cm X 12cm (3.1" x 4.7"), Warwick, RI Product Code: 0010211 The Ventrio Hernia Patch is a self-expanding non-absorbable... on . The reason FDA records is "Mislabeled: Product labeled as Small Oval (3.1" x 4.7") may contain Small Circle (3.0")". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1536-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Davol Inc., Sub. C.R. Bard, Inc.
Product
Bard Ventrio Small Oval Hernia Patch 8cm X 12cm (3.1" x 4.7"), Warwick, RI Product Code: 0010211 The Ventrio Hernia Patch is a self-expanding non-absorbable, sterile prosthesis, containing two primary layers of monofilament polypropylene mesh stitched with PTFE monofilament to an ePTFE sheet, forming a positioning pocket. The Ventrio Hernia Patch is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias.
Product code
FTL Mesh, Surgical, Polymeric
Reason for recall
Mislabeled: Product labeled as Small Oval (3.1" x 4.7") may contain Small Circle (3.0")
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
An "URGENT RECALL: BARD VENTRIO SMALL OVAL HERNIA PATCH (LOT DATB0015" letter dated April 8, 2010, was issued by the firm (Davol) to all customers via FedEx overnight with signature required. The letter describes the product, problem and action to be taken by the customers. The customers were instructed to: please examine their inventory and immediately discontinue use of this lot; contact the Davol Customer Service Department at 1-800-556-6275 for instructions on how to return product to firm; complete and fax the EFFECTIVENESS CHECK sheet to 1-401-825-8753, and forward a copy of this letter, if they are a distributor, to their customers who have received the product and lot number. All products returned will be held at the Davol, Warwick location. Please contact your Davol Territory Manager or C.R. Bard's Medical Service Support Department at 1-800-562-0027 if you have any other questions.
Quantity in commerce
198 units
Distribution
Nationwide distribution
Product codes and lots
REF 0010211 Lot DATB0015