BD Visitec EdgeAhead MVR Knife, BD Opthalmic Systems, Waltham, MA. Intended for use in cataract surgery to...
Becton Dickinson and Company began a recall of BD Visitec EdgeAhead MVR Knife, BD Opthalmic Systems, Waltham, MA. Intended for use in cataract surgery to create self-sealing stab incisions through the... on . The reason FDA records is "A field action was initiated due to a recent customer complaint from Japan there is a labeling mixup and that some of the BD Visitec EdgeAhead MVR Knife 0.90mm (20G), REF 585230, shelf packs may contain BD Visitec...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1597-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Visitec EdgeAhead MVR Knife, BD Opthalmic Systems, Waltham, MA. Intended for use in cataract surgery to create self-sealing stab incisions through the cornea or sclera. Also used in retinal surgery for sclerotomy incisions.
- Product code
- HNN Knife, Ophthalmic
- Reason for recall
- A field action was initiated due to a recent customer complaint from Japan there is a labeling mixup and that some of the BD Visitec EdgeAhead MVR Knife 0.90mm (20G), REF 585230, shelf packs may contain BD Visitec EdgeAhead Slit Knife 2.3mm (40 degree), REF 8009918.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- BD Medical issued a "Voluntary Advisory Notice" dated April 6, 2010. Consignees were informed of the affected product and asked to complete the enclosed Customer Response Form indicating whether the product will be returned to the firm. For further information, contact BD Medical Opthalmic Systems at 1-781-906-7952 or via email at [email protected].
- Quantity in commerce
- 1170 (US 810, OUS 360)
- Distribution
- Worldwide Distribution -- United States, Canada and Japan.
- Product codes and lots
- Model # 585230, Lot 9202221