Z-1325-2010
Pelton & Crane Dental Chairs, Spirit 2000 with magnetic headrest, Pelton & Crane, Charlotte, NC 28273...
KaVo Dental Technologies, LLC / initiated 2010-02-24 / Terminated / root cause: design
Pelton & Crane Company began a recall of Pelton & Crane Dental Chairs, Spirit 2000 with magnetic headrest, Pelton & Crane, Charlotte, NC 28273. Intended for the properly positioning of patients to... on . The reason FDA records is "Magnetic headrests may affect patients with implantable pacemakers and defibrillators". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1325-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- KaVo Dental Technologies, LLC
- Product
- Pelton & Crane Dental Chairs, Spirit 2000 with magnetic headrest, Pelton & Crane, Charlotte, NC 28273. Intended for the properly positioning of patients to perform dental procedures.
- Product code
- KLC Chair, Dental, With Operative Unit
- Reason for recall
- Magnetic headrests may affect patients with implantable pacemakers and defibrillators.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Pelton & Crane Issued an "Urgent Field Safety Alert" notification dated February 24, 2010. Consignees were instructed of the affected product and informed of proper actions to take. For further information, contact Pelton & Crane at 1-704-588-2126.
- Quantity in commerce
- Total: 498 chairs and 619 replacement upholstery spare parts
- Distribution
- Worldwide Distribution -- United States, Australia, the Netherlands, Canada and Germany.
- Product codes and lots
- All Devices with Model Number SP20