ViroMax viral/bacterial filter. The filter reduces the risk of infection due to cross-contamination between...
A M Systems Inc began a recall of ViroMax viral/bacterial filter. The filter reduces the risk of infection due to cross-contamination between patients For use on all patient populations in... on . The reason FDA records is "Some ViroMax viral/bacterial filters are leaking at the point where the two halves of the filter are ultrasonically bonded together". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0308-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- A-M Systems
- Product
- ViroMax viral/bacterial filter. The filter reduces the risk of infection due to cross-contamination between patients For use on all patient populations in conjunction with other respiratory devices containing standard 15mm and/or 22mm fittings (such as breathing circuits) to filter respiratory gases where infection from airborne bacteria and viruses is a concern. Disposable, for single patient use only.
- Product code
- CAH Filter, Bacterial, Breathing-Circuit
- Reason for recall
- Some ViroMax viral/bacterial filters are leaking at the point where the two halves of the filter are ultrasonically bonded together.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- A - M Systems sent a Urgent Notification of Product Recall/Correction dated March 31, 2010, to all affected customers. The letter identificed the product the problem and the action needed to be taken by the customer. Customers were instructed to cease use and/or distribution of the affected product and to immediately notify any of their sub-accounts with the same information. Customers are asked to provide A - M Systems with the amount distributed and amount remaining in inventory and to then destroy or discard any remaining product in their possession. Please call 360.683.8300, if you have any questions or concerns.
- Quantity in commerce
- 37800
- Distribution
- Worldwide Distribution -- USA ( nationwide) including the states of AZ, CA, FL, KY, MA, MI, NC, OH, OR, and WI., and the countries Canada, Mexico, and Taiwan.
- Product codes and lots
- Lot numbers: 551278, 551311, 551355, 551425, 551432, 551452, 551635, 551658, 551699, 551704, 551727, 551732, 551746, 551800, 551879, 551905, 551954 and 552095.