Z-1654-2010

Therakos CellEx Photopheresis System; Photopheresis Procedural Kits, for use with the THERAKOS CellEx...

Therakos Inc / initiated 2010-03-29 / Terminated / root cause: manufacturing process

Therakos Inc began a recall of Therakos CellEx Photopheresis System; Photopheresis Procedural Kits, for use with the THERAKOS CellEx Instrument on . The reason FDA records is "Higher than normal blood leaks on THERAKOS CELLEX Procedural Kits that are intended for use with the THERAKOS CELLEX Photopheresis System". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1654-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Therakos Inc
Product
Therakos CellEx Photopheresis System; Photopheresis Procedural Kits, for use with the THERAKOS CellEx Instrument;
Product code
LNR System, Photopheresis, Extracorporeal
Reason for recall
Higher than normal blood leaks on THERAKOS CELLEX Procedural Kits that are intended for use with the THERAKOS CELLEX Photopheresis System
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On March 29, 2010, Therakos Photopheresis sent Important Product Correction Notification letters by FedEx to all direct end use and distributor customers who received lot X120 in addition to phone contact. On March 31, 2010 those who received lot X123 were sent letters by Fed Ex as well as phone contact.. The one Canadian customer was notified by phone on March 31, 2010 followed by a Fed Ex letter. On April 12, consignees who were shipped Lot X119 were contacted by phone, followed by a hard copy letter. On May 7, 2010 all direct end use and distributor consignees were notified of the additional lot being added to the recall. Questions may be directed to Therakos Customer Technical Services at 1-877-865-6850.
Quantity in commerce
280 units in US (1 sales unit contains 3 kits each)
Distribution
Nationwide Distribution, and Canada and Germany
Product codes and lots
CLXUSA (US): Lot X120, Exp 12/10; X123, Exp 1/2011; Lot Y106, Exp 2/1/11. CLSECP (For European Union), Lot X119, Exp 12/2014; Lot X123; Exp 1/2015