Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1598-2010

ev3 Protg EverFlex Self-Expanding Biliary Stent System, Sterile EO, (REF: PRB35-06-150-080, REF...

Medtronic, Inc. / initiated 2010-03-31 / Terminated

Ev3, Inc. began a recall of ev3 Protg EverFlex Self-Expanding Biliary Stent System, Sterile EO, (REF: PRB35-06-150-080, REF PRB35-06-120-120 and PRB35-06-150-120), 4.5 - 5.5mm... on . The reason FDA records is "ev3 is conducting a voluntary recall on three lots of the Protg EverFlex" Biliary Stent System. They determined that a lot of 120mm Protg EverFlex" Biliary Stent system may contain a 150mm length Self-Expanding stent...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1598-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
ev3 Protg EverFlex Self-Expanding Biliary Stent System, Sterile EO, (REF: PRB35-06-150-080, REF PRB35-06-120-120 and PRB35-06-150-120), 4.5 - 5.5mm, Expiration: 2011-10-01, Rx only. ev3 Inc., 4500 Nathan Lane North, Plymouth, MN 55442-2929. Intended as a palliative treatment of malignant neoplasms in the biliary tree.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
ev3 is conducting a voluntary recall on three lots of the Protg EverFlex" Biliary Stent System. They determined that a lot of 120mm Protg EverFlex" Biliary Stent system may contain a 150mm length Self-Expanding stent and two lots o 150mm Protg EverFlex" Biliary Stent System may contain a 120 mm length Self-Expanding stent.
FDA root cause
Mixed-up of materials/components
Action
ev3 issued "Medical Device Recall" notifications dated March 31, 2010 addressed to "Dear Risk Manager or Cath Lab Manager". Each letter addressed the affected product. Users were instructed to locate and remove from use the affected product and a sales representative will be in contact shortly to arrange for return. For further information, contact ev3 Customer Service at 1-800-716-6700.
Quantity in commerce
19
Distribution
Nationwide Distribution -- CA, FL, IL, LA, MA, MI, NH, PA, TN, TX, UT, VA and WA.
Product codes and lots
Lot Numbers: 7857377, 7856788 and 7858698.