Z-1619-2010
Inserter Wand Holder, Non-Sterile, Serial # 4273-0000, Blackrock Microsystems, Salt Lake City, UT / or /...
Blackrock Microsystems / initiated 2010-03-26 / Terminated / root cause: design
I2s Micro Implantable Systems Llc began a recall of Inserter Wand Holder, Non-Sterile, Serial # 4273-0000, Blackrock Microsystems, Salt Lake City, UT / or / Cyberkinetics Neurotechnology Systems, Inc., Salt Lake... on . The reason FDA records is "Component of cranial array may not allow sufficient tightening to maintain inserter position". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1619-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Blackrock Microsystems
- Product
- Inserter Wand Holder, Non-Sterile, Serial # 4273-0000, Blackrock Microsystems, Salt Lake City, UT / or / Cyberkinetics Neurotechnology Systems, Inc., Salt Lake City, UT. Intended use: The Inserter Wand Holder serves two key functions. It firmly secures the Inserter Wand to prevent recoil during the NeuroPort Array insertion process, and it facilitates precision alignment of the Inserter Wand with respect to the back of the NeuroPort Array just prior to pneumatic impulse insertion.
- Product code
- GWL Amplifier, Physiological Signal
- Reason for recall
- Component of cranial array may not allow sufficient tightening to maintain inserter position.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, Blackrock Microsystems, sent a "RECALL" letter dated March 26, 2010, to all customers. The letter describes the product, problem and actions to be taken by the customers. The customers were instructed to discontinue use of all affected devices and promptly return any inventory, perform a physical account of your inventory for the recalled product and email a list of all serial numbers to Andrew Gotshalk at [email protected], and to please email Andrew Gotshalk even if you do not have any inserter wand holders or have destroyed or removed from use the inserter wand holders. Blackrock will rework or replace all Inserter Wand Holders that is returned under this recall. If you have any questions regarding this recall please call 801-582-5533 x 275.
- Quantity in commerce
- 22 units
- Distribution
- Nationwide distribution: CA, MA, NY, RI, and UT.
- Product codes and lots
- Serial # 4273-0001 through 4273-0061