Basic Core Pack # OR1805, Sterile, Tri State Hospital Supply Corporation Manufacturers of Centurion...
Centurion Medical Products began a recall of Basic Core Pack # OR1805, Sterile, Tri State Hospital Supply Corporation Manufacturers of Centurion Healthcare Products on . The reason FDA records is "The firm is conducting a sub recall of kits containing ConMed Goldline Rocker Switch Pencils. The pencils activation switch may remain in the "on" position after the switch has been depressed with excessive force...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1595-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Centurion Medical Products, LP
- Product
- Basic Core Pack # OR1805, Sterile, Tri State Hospital Supply Corporation Manufacturers of Centurion Healthcare Products.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- The firm is conducting a sub recall of kits containing ConMed Goldline Rocker Switch Pencils. The pencils activation switch may remain in the "on" position after the switch has been depressed with excessive force, thereby continuously activating the pencil.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Centurion sent out an "Urgent Product Recall Notice" to one customer on 4/9/2010. The customer was instructed to discontinue use of the product. The customer was also instructed to return any instock product .
- Quantity in commerce
- 74 kits
- Distribution
- TX
- Product codes and lots
- 2009100590, 2009110990, 2009121490, 2010011190 and 2010031590.
FIRST ARTICLE / WEEKLY