Z-1469-2010
Shiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 10 FEN, Assembled in Mexico, Tyco Healthcare Group...
Medtronic, Inc. / initiated 2010-04-13 / Terminated / root cause: manufacturing process
Nellcor Puritan Bennett Inc. (dba Covidien Ltd) began a recall of Shiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 10 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton... on . The reason FDA records is "Inflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1469-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Shiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 10 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper airway obstructions, provide long term ventilation, support and/or manage tracheal/bronchial secretions.
- Product code
- JOH Tube Tracheostomy And Tube Cuff
- Reason for recall
- Inflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Covidien issued an "Urgent Medical Device Recall" notification dated April 13, 2010. Consignees were alerted to the affected device. If an affected product is already in use, the firm recommends that the tube be replaced as soon as the patient's physician determines it is clinically appropriate. All affected product is to be returned to the firm. For further Information, contact Covidien at 1-800-635-5267.
- Quantity in commerce
- 3,912 units
- Distribution
- Worldwide Distribution -- United States (including Guam and Puerto Rico), Australia, Belgium, Brazil, Canada, Chile, Colombia, Honduras, Japan, Mexico, Panama, Singapore and Uruguay.
- Product codes and lots
- Lot Numbers: 0811001364 through 0908000233