Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1527-2010

GE, 1009-9002-000 Avance Anesthesia Machine and Monitor

GE Healthcare / initiated 2010-03-12 / Terminated / root cause: manufacturing process

Ge Healthcare, Llc began a recall of GE, 1009-9002-000 Avance Anesthesia Machine and Monitor on . The reason FDA records is "Unanticipated shut-down: A specific lot of GE Healthcare Aisys and Avance machines control board wiring harness have a defect, which can cause the machine to unexpectedly shut down, terminating ventilation, anesthetic...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1527-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE, 1009-9002-000 Avance Anesthesia Machine and Monitor
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
Unanticipated shut-down: A specific lot of GE Healthcare Aisys and Avance machines control board wiring harness have a defect, which can cause the machine to unexpectedly shut down, terminating ventilation, anesthetic delivery, and potentially patient monitoring.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated March 12, 2010. The letter was addressed to Chief of Anesthesia, Health Care Administrator/Risk Manager, Director of Biomedical/Clinical Engineering. The letter listed the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Quantity in commerce
18 (4 USA, 14 OUS)
Distribution
Worldwide distribution: USA (CA, FL, TN), CANADA, DENMARK, FINLAND, FRANCE, ITALY, SAUDI ARABIA, SOUTH AFRICA, SPAIN, UNITED ARAB EMIRATES, and UNITED KINGDOM.
Product codes and lots
Serial Numbers: ANBN01180, ANBN01187, ANBN01200, ANBN01201, ANBN01211, ANBN01212, ANBN01213, ANBN01220, ANBN01223, ANBN01224, ANBN01250, ANBN01266, ANBN01284, ANBN01285, ANBN01286, ANBN01287, ANBN01288, ANBN01289