GE, 1009-9002-000 Avance Anesthesia Machine and Monitor
Ge Healthcare, Llc began a recall of GE, 1009-9002-000 Avance Anesthesia Machine and Monitor on . The reason FDA records is "Unanticipated shut-down: A specific lot of GE Healthcare Aisys and Avance machines control board wiring harness have a defect, which can cause the machine to unexpectedly shut down, terminating ventilation, anesthetic...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1527-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE, 1009-9002-000 Avance Anesthesia Machine and Monitor
- Product code
- BSZ Gas-Machine, Anesthesia
- Reason for recall
- Unanticipated shut-down: A specific lot of GE Healthcare Aisys and Avance machines control board wiring harness have a defect, which can cause the machine to unexpectedly shut down, terminating ventilation, anesthetic delivery, and potentially patient monitoring.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated March 12, 2010. The letter was addressed to Chief of Anesthesia, Health Care Administrator/Risk Manager, Director of Biomedical/Clinical Engineering. The letter listed the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
- Quantity in commerce
- 18 (4 USA, 14 OUS)
- Distribution
- Worldwide distribution: USA (CA, FL, TN), CANADA, DENMARK, FINLAND, FRANCE, ITALY, SAUDI ARABIA, SOUTH AFRICA, SPAIN, UNITED ARAB EMIRATES, and UNITED KINGDOM.
- Product codes and lots
- Serial Numbers: ANBN01180, ANBN01187, ANBN01200, ANBN01201, ANBN01211, ANBN01212, ANBN01213, ANBN01220, ANBN01223, ANBN01224, ANBN01250, ANBN01266, ANBN01284, ANBN01285, ANBN01286, ANBN01287, ANBN01288, ANBN01289