Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1491-2010

Shiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 10 LPC-S, Assembled in Mexico, Tyco...

Medtronic, Inc. / initiated 2010-04-13 / Terminated / root cause: manufacturing process

Nellcor Puritan Bennett Inc. (dba Covidien Ltd) began a recall of Shiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 10 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division... on . The reason FDA records is "Inflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1491-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Shiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 10 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper airway obstructions, provide long term ventilation, support and/or manage tracheal/bronchial secretions.
Product code
JOH Tube Tracheostomy And Tube Cuff
Reason for recall
Inflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Covidien issued an "Urgent Medical Device Recall" notification dated April 13, 2010. Consignees were alerted to the affected device. If an affected product is already in use, the firm recommends that the tube be replaced as soon as the patient's physician determines it is clinically appropriate. All affected product is to be returned to the firm. For further Information, contact Covidien at 1-800-635-5267.
Quantity in commerce
223 units
Distribution
Worldwide Distribution -- United States (including Guam and Puerto Rico), Australia, Belgium, Brazil, Canada, Chile, Colombia, Honduras, Japan, Mexico, Panama, Singapore and Uruguay.
Product codes and lots
Lot Numbers: 0812000912 through 0907000367