Z-2482-2010

ARCHITECT i1000SR" Processing Module System For In Vitro Diagnostic Use, Abbott Laboratories Abbott Park, IL...

Abbott Medical / initiated 2008-10-21 / Terminated / root cause: manufacturing process

Abbott Laboratories, Inc began a recall of ARCHITECT i1000SR" Processing Module System For In Vitro Diagnostic Use, Abbott Laboratories Abbott Park, IL Product List Number, Part # or Model: 01L86-01 In... on . The reason FDA records is "During manufacturing, two components of the ARCHITECT i1000SR were reversed which could potentially cause falsely elevated patient results due to a sample-reagent and reagent-sample carryover". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2482-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ARCHITECT i1000SR" Processing Module System For In Vitro Diagnostic Use, Abbott Laboratories Abbott Park, IL Product List Number, Part # or Model: 01L86-01 In Vitro Diagnostic use only. The Abbott ARCHITECT System is designed to perform automated: Chemistry tests, utilizing photometry and potentiometric technology, and Immunoassay tests, utilizing CMIA (chemiluminescent microparticle assay) detection technology.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
During manufacturing, two components of the ARCHITECT i1000SR were reversed which could potentially cause falsely elevated patient results due to a sample-reagent and reagent-sample carryover.
Root cause tag
manufacturing process
FDA root cause
Mixed-up of materials/components
Action
ABBOTT issued Technical Service Bulletins to Field Service Representatives who then inspected every unit .
Quantity in commerce
404 units distributed worldwide (including US)
Distribution
Worldwide distribution:USA including states of: CO, DC, FL, GA, IL, LA, MA, MN, NY, OH, OK, TX, UT, and WV and countries including: Australia, Brazil, Bogota, Canada, China WFOTC, Columbia, Curacao (NETH ANT), Germany, Hong Kong, Japan, Mexico, Singapore, and Taiwan.
Product codes and lots
Serial Numbers: (1st recall) i1SR01001-1224; 1226-1300; 1302-1306; 1308-1313; 1315-1316; 1318-1352; 1355-1363; 1365; 1376; 1381-1382  Extended to include serial numbers manufactured from 04/01/08 to 03/31/09: i1SR01021, 1026, 1032, 1040, 1036, 1043, 1059, 1065, 1069, 1071, 1071, 1074, 1082, 1094, 1106, 1107, 1109, 1111, 1112, 1114, 1117, 1119, 1121, 1123, 1128, 1133, 1135, 1138, 1139, 1158, 1162, 1167, 1171, 1173, 1177, 1183, 1186, 1188, 1189, 1198, 1200, 1201, 1205, 1209, 1211, 1217, 1225, 1230, 1243, 1246, 1249, 1250, 1260, 1262, 1272, 1277, 1279, 1284, 1286-1288, 1290, 1294, 1301, 1304, 1312, 1317, 1321, 1324-1327, 1332-1335, 1338, 1339, 1342-1345, 1347, 1348, 1351-1356, 1363-1368, 1370-1372, 1375, 1377-1387, 1389-1396, 1398-1410, 1412-1414, 1416-1441, 1443-1530, 1532-1536 (some of these #'s were removed whose field status was Demo/Training and instruments that had TSB 117-004 completed prior to Feb 19.