Z-2216-2010
Trabecular Metal acetabular revision system column buttress left posterior/right anterior, 00-4894-300-00...
Zimmer GmbH / initiated 2010-04-06 / Terminated / root cause: packaging
Zimmer Inc. began a recall of Trabecular Metal acetabular revision system column buttress left posterior/right anterior, 00-4894-300-00, sterile, Zimmer Inc., Parsippany, NJ. Usage... on . The reason FDA records is "The firm has determined that some of the packages could develop a breach of the seal in the sterile barrier pouches". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2216-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Trabecular Metal acetabular revision system column buttress left posterior/right anterior, 00-4894-300-00, sterile, Zimmer Inc., Parsippany, NJ. Usage: Implantable devices that are used to address bony deficiencies of the pelvic antebulum in revision total hip arthroplasty surgery.
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- The firm has determined that some of the packages could develop a breach of the seal in the sterile barrier pouches.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- The firm sent an "Urgent Device Removal" letter dated 4/6/2010. The consignees were instructed to inventory the products on hand and return the recalled product to the manufacturer. More information is available at 1-800-613-6131.
- Quantity in commerce
- 1,972 all products
- Distribution
- US, Germany, Brazil, Singapore, Canada, Australia, England, India, and Switzerland.
- Product codes and lots
- 60492196, 60531086, 60703467, 60667441, 60728123, 60767106, 60787931, 60808260, 60904605, 61024439, 61143012, 61217116 and 61355988.