Z-2216-2010

Trabecular Metal acetabular revision system column buttress left posterior/right anterior, 00-4894-300-00...

Zimmer GmbH / initiated 2010-04-06 / Terminated / root cause: packaging

Zimmer Inc. began a recall of Trabecular Metal acetabular revision system column buttress left posterior/right anterior, 00-4894-300-00, sterile, Zimmer Inc., Parsippany, NJ. Usage... on . The reason FDA records is "The firm has determined that some of the packages could develop a breach of the seal in the sterile barrier pouches". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2216-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Trabecular Metal acetabular revision system column buttress left posterior/right anterior, 00-4894-300-00, sterile, Zimmer Inc., Parsippany, NJ. Usage: Implantable devices that are used to address bony deficiencies of the pelvic antebulum in revision total hip arthroplasty surgery.
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
The firm has determined that some of the packages could develop a breach of the seal in the sterile barrier pouches.
Root cause tag
packaging
FDA root cause
Packaging
Action
The firm sent an "Urgent Device Removal" letter dated 4/6/2010. The consignees were instructed to inventory the products on hand and return the recalled product to the manufacturer. More information is available at 1-800-613-6131.
Quantity in commerce
1,972 all products
Distribution
US, Germany, Brazil, Singapore, Canada, Australia, England, India, and Switzerland.
Product codes and lots
60492196, 60531086, 60703467, 60667441, 60728123, 60767106, 60787931, 60808260, 60904605, 61024439, 61143012, 61217116 and 61355988.