Z-1667-2010
Mallory Head Calcar Femoral, 34 mm resection/porous coated, 9mm x 170 mm stem, Sterile, REF 11-104902, Biomet...
Zimmer GmbH / initiated 2010-03-10 / Terminated
Biomet, Inc. began a recall of Mallory Head Calcar Femoral, 34 mm resection/porous coated, 9mm x 170 mm stem, Sterile, REF 11-104902, Biomet Orthopaedics, Warsaw, IN. Implant device intended... on . The reason FDA records is "The firm has found that some lots of Mallory Head Revision Stems had flute geometry that was out of specification. The flute geometry was too deep". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1667-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Mallory Head Calcar Femoral, 34 mm resection/porous coated, 9mm x 170 mm stem, Sterile, REF 11-104902, Biomet Orthopaedics, Warsaw, IN. Implant device intended for use in femoral revision surgery.
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- The firm has found that some lots of Mallory Head Revision Stems had flute geometry that was out of specification. The flute geometry was too deep.
- FDA root cause
- Process change control
- Action
- Biomet Orthopedics issued an "Urgent Medical Device Recall Notice" dated March 10, 2010. The letter went to distributors and users of the recalled device. The letter instructed the consignees to locate the recalled product and return the product to the firm. For further information, contact Biomet at 1-800-348-9500 or 1-574-372-3983, Monday through Friday, 8:00 AM to 5:00 PM.
- Quantity in commerce
- 10
- Distribution
- Worldwide Distribution -- United States (OH, IN, FL, KY, NV and CA), Mexico, Chile, China, Japan and The Netherlands.
- Product codes and lots
- Lot Numbers: M650320, M298660.