Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2120-2010

Howmedica Osteonics Corp., Stryker Orthopaedics Passport A.R. Knee Instrumentation Anterior Skim Guide...

Howmedica Osteonics Corp / initiated 2010-04-05 / Terminated / root cause: design

Stryker Howmedica Osteonics Corp. began a recall of Howmedica Osteonics Corp., Stryker Orthopaedics Passport A.R. Knee Instrumentation Anterior Skim Guide; Catalog Number 7650-5003 on . The reason FDA records is "One of the two posts on the Anterior Resection Guide could disassociate from the main body of the part which has been determined to be caused by a fracture weld". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2120-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Howmedica Osteonics Corp., Stryker Orthopaedics Passport A.R. Knee Instrumentation Anterior Skim Guide; Catalog Number 7650-5003.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
One of the two posts on the Anterior Resection Guide could disassociate from the main body of the part which has been determined to be caused by a fracture weld.
Root cause tag
design
FDA root cause
Device Design
Action
Stryker Urgent Product Recall letters dated April 5, 2010, were sent out via Federal Express to Stryker branches/agencies, Hospital Risk Management, hospital Chief of Orthopedics and Surgeons. All hospitals and surgeons that used the product in the last year were notified. Questions and/or concerns should be directed to Rich Wolyn at 201-972-2100.
Quantity in commerce
682 units in US, 1292 units internationally.
Distribution
Worldwide distribution: USA, Ireland, Germany, Sweden, The Netherlands, France, Italy, Switzerland, South Africa, United Kingdom, Japan, Argentina, Mexico, Miramar (city), China, Taiwan, India, Australia, Korea, New Zealand, Malaysia, Singapore, and Canada.
Product codes and lots
Catalog Number 7650-5003; all lots.