Integrated Data Management System - Version 7.1, 7.2 and 7.3; ITC Corporation, 8 Olsen Place South...
International Technidyne Corporation began a recall of Integrated Data Management System - Version 7.1, 7.2 and 7.3; ITC Corporation, 8 Olsen Place South, Piscataway, NJ 08854 on . The reason FDA records is "A tabulation error in the Integrated Data Management Systems (IDMS) data management software that results in incorrect Electronic Quality Control (EQC) Summary Reports: The Operator and the Device Performance Summary...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2438-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Accriva Diagnostics, Inc.
- Product
- Integrated Data Management System - Version 7.1, 7.2 and 7.3; ITC Corporation, 8 Olsen Place South, Piscataway, NJ 08854
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- A tabulation error in the Integrated Data Management Systems (IDMS) data management software that results in incorrect Electronic Quality Control (EQC) Summary Reports: The Operator and the Device Performance Summary Reports. The IDMS software is not used for the clinical management of individual patients.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- ITC sent an URGENT MEDICAL DEVICE CORRECTION letter dated March 10, 2010 to Distributors and First Consignee End Users identifying the affected product and affected summary reports, the reason for recall, and the actions to be taken by the distributors and customers. New Errata Instructions were enclosed with each letter and are to be put in the idms user guide for reference. The new instructions are to be followed when running reports for the ESV or TVT tests run by the Hemochron Response. The enclosed Customer Account Tracking Form is to be completed and returned to ITC indicating that the letter has been received and appropriate questions answered as to the use of the software system. Distributors are asked to check inventory for recalled product. If affected product is found, shipment should be stopped and placed on hold. Distributors should notify customers of the correction and provide them with a copy of the correction letter and updated instructions. This should be done by traceable mail. Distributors should return the account tracking form and return affected software disks for replacement. Questions should be directed to ITC Technical Support at 800-631-5945.
- Quantity in commerce
- 84 total idms software systems; 66 in US, 18 outside of US.
- Distribution
- Worldwide distribution: USA including the states of AL, AZ, CA, FL, GA, IL, KS, LA, MD, MI, NC, ND, NJ, NM, NY, OH, PA, TN , TX, UT, VA, WA, and WI, and the countries of Italy, Turkey, Canada, France, Belgium, Germany, UAE, and New Zealand.
- Product codes and lots
- Product code 464100, Version numbers 7.1.0, 7.2.0 and 7.3.0.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
