Mycobacterium Avium Complex Culture Identification Test Identification Test Kit, 2835
Gen-Probe Inc began a recall of Mycobacterium Avium Complex Culture Identification Test Identification Test Kit, 2835 on . The reason FDA records is "This recall has been initiated following the identification of certain assay tubes that do not contain probe reagent. The affected kits may contain a combination of tubes with correct probe amount, partial probe amount...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1437-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hologic, Inc.
- Product
- Mycobacterium Avium Complex Culture Identification Test Identification Test Kit, 2835.
- Product code
- LQF Dna-Reagents, Mycobacterium Spp.
- Reason for recall
- This recall has been initiated following the identification of certain assay tubes that do not contain probe reagent. The affected kits may contain a combination of tubes with correct probe amount, partial probe amount and some probe tubes that are empty. Possible false negative results.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Gen-Probe issued a Recall Notice to customers recommending that they discontinue the use of the identified batch (lot) numbers listed. For customers that have used these batches and obtained negative results, they recommend that the referring physician be notified about the possibility of a false negative result. With the Recall Notice, consignees received a Customer Response Form to complete. Gen-Probe requested that they segregate the affected batches listed on the Customer Response Form, then reconcile, and destroy all partial and/or full kits left in inventory, being sure to document the amount destroyed on the Customer Response Form. Gen-Probe should be updated via phone, fax, and email information. Customers should then sign and return the completed form to Gen-Probe via fax to 858-410-8250, or scanned and emailed to [email protected]. The Technical Support team should be contacted for replacement kits. Technical Support was also available for additional information at 888-484-4747 or [email protected].
- Quantity in commerce
- 311 kits
- Distribution
- Worldwide Distribution -- USA, including Puerto Rico and countries of Australia, Guatemala, Malaysia, Bahamas, Hong Kong, Mali, Singapore, Bermuda, India, Mexico, South Africa, Canada, Israel, Netherlands, South Korea, Finland, Kuwait, Panama, Thailand, France, Latvia, and Philippines.
- Product codes and lots
- Lots (batches): 555767