Z-1636-2010

ADVANCE DURAMER TIBIAL INSERT, REF: KIMP-312L, SIZE 3 LEFT MEDIAL PIVOT, THICKNESS 12MM, 1 EACH, STERILE EO...

Wright Medical Technology Inc / initiated 2009-11-24 / Terminated / root cause: packaging

Wright Medical Technology Inc began a recall of ADVANCE DURAMER TIBIAL INSERT, REF: KIMP-312L, SIZE 3 LEFT MEDIAL PIVOT, THICKNESS 12MM, 1 EACH, STERILE EO, Wright Medical Technology, 4677 Airline Road... on . The reason FDA records is "The left medial pivot and right medial pivot inserts were co-mingled". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1636-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Wright Medical Technology Inc
Product
ADVANCE DURAMER TIBIAL INSERT, REF: KIMP-312L, SIZE 3 LEFT MEDIAL PIVOT, THICKNESS 12MM, 1 EACH, STERILE EO, Wright Medical Technology, 4677 Airline Road, Arlington, TN 38002. Indicated for use in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The left medial pivot and right medial pivot inserts were co-mingled
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Wright Medical Technology, Inc. issued an "Urgent: Medical Device Voluntary Recall" notification dated November 24, 2009. The letters were addressed to distributors, hospital administrators, and surgeons. Consignees were asked to return all affected product in addition to a completed Response Form to the firm. For further information, contact Wright Medical Technology, Inc. at 1-800-874-5630.
Quantity in commerce
23 units
Distribution
Worldwide Distribution -- United States, The Netherlands, Italy, Republic of Korea and Japan.
Product codes and lots
Lot Number: 679391