Z-1804-2010

Synthes Medullary Tubes

Synthes GmbH / initiated 2010-04-01 / Terminated / root cause: sterility

Synthes USA (HQ), Inc. began a recall of Synthes Medullary Tubes on . The reason FDA records is "After repeated sterilizations product may become discolored and break including while in the medullary canal. Synthes has determined that the Medullary Tube will be a single use item that is to be discarded after one...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-1804-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes Medullary Tubes
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
After repeated sterilizations product may become discolored and break including while in the medullary canal. Synthes has determined that the Medullary Tube will be a single use item that is to be discarded after one sterilization cycle.
Root cause tag
sterility
FDA root cause
Other
Action
The recalling firm issued an Urgent medical Device Recall letter dated 4/9/10 to all trauma hospital accounts informing them of the problem and the need to return the products. All sales consultants were notified via email and letter. Questions are directed to the company at 800-620-7025 x 5452 or 610-719-5452, or contact your Synthes Trauma Sales Consultant.
Quantity in commerce
18,505 tubes
Distribution
The product was shipped to medical facilities nationwide.
Product codes and lots
Catalog number 355.01. all lot numbers