Z-1804-2010
Synthes Medullary Tubes
Synthes GmbH / initiated 2010-04-01 / Terminated / root cause: sterility
Synthes USA (HQ), Inc. began a recall of Synthes Medullary Tubes on . The reason FDA records is "After repeated sterilizations product may become discolored and break including while in the medullary canal. Synthes has determined that the Medullary Tube will be a single use item that is to be discarded after one...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1804-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes Medullary Tubes
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- After repeated sterilizations product may become discolored and break including while in the medullary canal. Synthes has determined that the Medullary Tube will be a single use item that is to be discarded after one sterilization cycle.
- Root cause tag
- sterility
- FDA root cause
- Other
- Action
- The recalling firm issued an Urgent medical Device Recall letter dated 4/9/10 to all trauma hospital accounts informing them of the problem and the need to return the products. All sales consultants were notified via email and letter. Questions are directed to the company at 800-620-7025 x 5452 or 610-719-5452, or contact your Synthes Trauma Sales Consultant.
- Quantity in commerce
- 18,505 tubes
- Distribution
- The product was shipped to medical facilities nationwide.
- Product codes and lots
- Catalog number 355.01. all lot numbers