Z-1591-2010

VNS Therapy Demipulse Generator, Model 103 and VNS Therapy Demipulse Duo Generator, Model 104 VNS Therapy...

LivaNova USA, Inc. / initiated 2010-04-06 / Terminated / root cause: software

Cyberonics, Inc began a recall of VNS Therapy Demipulse Generator, Model 103 and VNS Therapy Demipulse Duo Generator, Model 104 VNS Therapy Programming System, Model 250 Version 7.0 and 7.1... on . The reason FDA records is "Battery life projection is inaccurate". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1591-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
VNS Therapy Demipulse Generator, Model 103 and VNS Therapy Demipulse Duo Generator, Model 104 VNS Therapy Programming System, Model 250 Version 7.0 and 7.1 Software
Product code
LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
Reason for recall
Battery life projection is inaccurate.
Root cause tag
software
FDA root cause
Software design
Action
Safety Alert sent April 6, 2010 to all physicians of record for Model 103 or Model 104 generators and to all hospitals with non-implanted affected product, that explains the event, risks presented and recommended actions to take. The generator can be reprogrammed by the physician the next time the patient is seen in the clinic. The issue can also be corrected during the initial implant procedure with the instructions provided. Additional questions should be directed to the firm; United States customers can contact Clinical Technical Support at 866-882-8804.
Quantity in commerce
Model103: 8109 units; Model 104: 1722 units; Programmers: 8891
Distribution
Worldwide distribution.
Product codes and lots
Model 103, VNS Therapy Demipulse Generator Model 104, VNS Therapy Demipulse Duo Generator Model 250, Programming System (version 7.0 or 7.1)