SEQUENCER for Mosaiq interfaced to Electa SL75-5 machines. Sequencer Verify and Record System is to assist in...
Impac Medical Systems Inc began a recall of SEQUENCER for Mosaiq interfaced to Electa SL75-5 machines. Sequencer Verify and Record System is to assist in the process of patient treatment on a... on . The reason FDA records is "Computer interface is defected. Computer based verification system for recording patient treatment on a radiotherapy treatment machine fails to record treated field". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1029-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- SEQUENCER for Mosaiq interfaced to Electa SL75-5 machines. Sequencer Verify and Record System is to assist in the process of patient treatment on a radiotherapy treatment machine with software control.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Computer interface is defected. Computer based verification system for recording patient treatment on a radiotherapy treatment machine fails to record treated field.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- A User Notice was sent by the firm to its consignees. It identified the affected product and the reason for the recall. It also discussed the clinical impact and interim mitigations. The firm states in the letter that a fix for the problem and been developed and is available. Customers are to contact IMPAC Support for the patch to the SL75-5 interface. If customers have questions, concerns, or requests, they can email [email protected] or one of the numbers listed in the letter for the applicable country. The letter should be distributed to all users affected by the issue at the customer's organization.
- Quantity in commerce
- 40
- Distribution
- Worldwide Distribution -- US, Canada, Germany, and France.
- Product codes and lots
- Sequencer for MOSAIQ versions: 1.20, 1.30 ; 1.50 , 1.60 and 2.00, interfacing to an Elekta SL 75-5 Linacs.