UniCel DxH 800 Coulter Cellular Analysis System, Part Number: 629029. Beckman Coulter, Inc. Brea, CA...
Beckman Coulter Inc. began a recall of UniCel DxH 800 Coulter Cellular Analysis System, Part Number: 629029. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer for in vitro... on . The reason FDA records is "The Corrective Action was initiated after Beckman Coulter confirmed three issues related to host transmissions with the UniCel DxH 800. This information only affects instruments interfaced to a host system". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1680-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- UniCel DxH 800 Coulter Cellular Analysis System, Part Number: 629029. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- The Corrective Action was initiated after Beckman Coulter confirmed three issues related to host transmissions with the UniCel DxH 800. This information only affects instruments interfaced to a host system.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Beckman Coulter, Inc. issued an "Urgent: Product Corrective Action" notification dated December 8, 2009. Consignees were informed of the affected product and provided temporary solutions until further support will be made available. For additional information, contact your local Beckman Coulter Representative or call 800-526-7694.
- Quantity in commerce
- 42 units (40 in US)
- Distribution
- Distributed Nationwide and Canada.
- Product codes and lots
- All Serial Numbers