Z-1680-2010

UniCel DxH 800 Coulter Cellular Analysis System, Part Number: 629029. Beckman Coulter, Inc. Brea, CA...

Siemens Healthcare Diagnostics Inc. / initiated 2009-12-08 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of UniCel DxH 800 Coulter Cellular Analysis System, Part Number: 629029. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer for in vitro... on . The reason FDA records is "The Corrective Action was initiated after Beckman Coulter confirmed three issues related to host transmissions with the UniCel DxH 800. This information only affects instruments interfaced to a host system". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1680-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
UniCel DxH 800 Coulter Cellular Analysis System, Part Number: 629029. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.
Product code
GKZ Counter, Differential Cell
Reason for recall
The Corrective Action was initiated after Beckman Coulter confirmed three issues related to host transmissions with the UniCel DxH 800. This information only affects instruments interfaced to a host system.
Root cause tag
software
FDA root cause
Software design
Action
Beckman Coulter, Inc. issued an "Urgent: Product Corrective Action" notification dated December 8, 2009. Consignees were informed of the affected product and provided temporary solutions until further support will be made available. For additional information, contact your local Beckman Coulter Representative or call 800-526-7694.
Quantity in commerce
42 units (40 in US)
Distribution
Distributed Nationwide and Canada.
Product codes and lots
All Serial Numbers