Z-1438-2011

Mycobacterium Tuberculosis Complex Culture Identification Test, 2860

Hologic, Inc. / initiated 2008-12-30 / Terminated / root cause: manufacturing process

Gen-Probe Inc began a recall of Mycobacterium Tuberculosis Complex Culture Identification Test, 2860 on . The reason FDA records is "This recall has been initiated following the identification of certain assay tubes that do not contain probe reagent. The affected kits may contain a combination of tubes with correct probe amount, partial probe amount...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1438-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hologic, Inc.
Product
Mycobacterium Tuberculosis Complex Culture Identification Test, 2860.
Product code
LQF Dna-Reagents, Mycobacterium Spp.
Reason for recall
This recall has been initiated following the identification of certain assay tubes that do not contain probe reagent. The affected kits may contain a combination of tubes with correct probe amount, partial probe amount and some probe tubes that are empty. Possible false negative results.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Gen-Probe issued a Recall Notice to customers recommending that they discontinue the use of the identified batch (lot) numbers listed. For customers that have used these batches and obtained negative results, they recommend that the referring physician be notified about the possibility of a false negative result. With the Recall Notice, consignees received a Customer Response Form to complete. Gen-Probe requested that they segregate the affected batches listed on the Customer Response Form, then reconcile, and destroy all partial and/or full kits left in inventory, being sure to document the amount destroyed on the Customer Response Form. Gen-Probe should be updated via phone, fax, and email information. Customers should then sign and return the completed form to Gen-Probe via fax to 858-410-8250, or scanned and emailed to [email protected]. The Technical Support team should be contacted for replacement kits. Technical Support was also available for additional information at 888-484-4747 or [email protected].
Quantity in commerce
1293 kits
Distribution
Worldwide Distribution -- USA, including Puerto Rico and countries of Australia, Guatemala, Malaysia, Bahamas, Hong Kong, Mali, Singapore, Bermuda, India, Mexico, South Africa, Canada, Israel, Netherlands, South Korea, Finland, Kuwait, Panama, Thailand, France, Latvia, and Philippines.
Product codes and lots
Lots (batches): 554628, 555170, 556204 (Foreign Lot): 554840