Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1599-2010

Eon Mini Implantable Pulse Generator (IPG). Advanced Neuromodulation Systems, Inc. Plano, TX. Indicated as an...

Abbott Medical / initiated 2010-04-23 / Terminated / root cause: manufacturing process

Advanced Neuromodulation Systems, Inc began a recall of Eon Mini Implantable Pulse Generator (IPG). Advanced Neuromodulation Systems, Inc. Plano, TX. Indicated as an aid in the management of chronic, intractable... on . The reason FDA records is "Due to a missed inspection step during manufacturing, a small number of devices may experience a gradual or sudden loss of charging capability". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1599-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Eon Mini Implantable Pulse Generator (IPG). Advanced Neuromodulation Systems, Inc. Plano, TX. Indicated as an aid in the management of chronic, intractable pain of the trunk and/or limbs, including unilateral or bilateral pain.
Product code
LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Reason for recall
Due to a missed inspection step during manufacturing, a small number of devices may experience a gradual or sudden loss of charging capability.
Root cause tag
manufacturing process
FDA root cause
Manufacturing material removal
Action
St. Jude Medical issued an "Important Medical Device Information" notification dated April 2010. Consignees were informed of the affected product and instructed on necessary steps. Any returned product should be sent to the firm. For further information, contact your St. Jude Medical Neuromodulation Division Sales Representative or call 1-972-309-2154.
Quantity in commerce
7 devices
Distribution
Nationwide Distribution -- GA, NC, CA, OK, SC and NJ.
Product codes and lots
Device Code: 65-3788-01, Serial Numbers: 703271, 703287, 703050, 703051, 704169, 704476 and 704520.