Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1611-2010

4.5MM Bone Tap, Modular; Manufactured by: Blackstone Medical, Inc. 1720 Bray Central Drive, McKinney, TX...

Blackstone Medical, Inc. / initiated 2010-04-21 / Terminated / root cause: manufacturing process

Blackstone Medical, Inc. began a recall of 4.5MM Bone Tap, Modular; Manufactured by: Blackstone Medical, Inc. 1720 Bray Central Drive, McKinney, TX 75069. Intended to created a screw thread in a hole... on . The reason FDA records is "Certain lots of the non-cannulated Modular Bone Taps used with the Firebird Spinal Fixation Systems may contain a manufacturing non conformance resulting in the taps creating a single lead screw thread rather than the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1611-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Blackstone Medical, Inc.
Product
4.5MM Bone Tap, Modular; Manufactured by: Blackstone Medical, Inc. 1720 Bray Central Drive, McKinney, TX 75069. Intended to created a screw thread in a hole drilled into bone to facilitate placement of a bone screw.
Product code
HWX Tap, Bone
Reason for recall
Certain lots of the non-cannulated Modular Bone Taps used with the Firebird Spinal Fixation Systems may contain a manufacturing non conformance resulting in the taps creating a single lead screw thread rather than the required dual lead screw thread.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Orthofix issued an "Urgent Product Recall" notification dated April 21, 2010 by USPS certified mail to hospitals and distributors. Consignees were asked to return all affected product to the firm. For additional information, contact the Orthofix Regulatory Affairs Department at 1-973-406-2847.
Quantity in commerce
89
Distribution
Nationwide and to one location each in Australia and Spain.
Product codes and lots
Part Number 52-1024; Lot numbers16380-xxxx and 17162-xxxx