Z-0864-2011

Pioneer Plus Catheter PPlus 120 dual lumen device, inserted through a 6F introducer sheath and placed...

Medtronic, Inc. / initiated 2010-04-30 / Terminated / root cause: manufacturing process

Medtronic CardioVascular began a recall of Pioneer Plus Catheter PPlus 120 dual lumen device, inserted through a 6F introducer sheath and placed percutaneously into a peripheral vessel. The device... on . The reason FDA records is "Needles may fail to retract and thus cause injury". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0864-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Pioneer Plus Catheter PPlus 120 dual lumen device, inserted through a 6F introducer sheath and placed percutaneously into a peripheral vessel. The device tracks to its intended site in the vasculature over a standard length, commercially available 0.014" (0.36 mm) Rapid Exchange (RX) tracking guide wire. It utilizes an extendable, hollow Nitinol guide tip (needle) to facilitate the redirection and placement of a 0.014 (0.36 mm) Over the Wire (OTW) guide wire into peripheral vessels. Specification Developer: Medtronic, Minneapolis, MN. Manufactured by: Accellent, Laconia NH The Pioneer Plus Catheter is intended to facilitate placement and positioning of catheters within the peripheral vasculature. The Pioneer Plus Catheter also provides an intra-luminal cross-sectional ultrasound image of the area of interest to facilitate placement of guide wires beyond stenotic lesions (e.g., sub-total, total or chronic total occlusions) prior to additional intervention (i.e. PTCA, stent, etc.). The Pioneer Plus Catheter is not indicated for use in the coronary or cerebral vasculature.
Product code
DQY Catheter, Percutaneous
Reason for recall
Needles may fail to retract and thus cause injury.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Medtronic sent an Urgent Medical Device Safety Information letter to all affected customers. The letter identified the product, the problem,and the action to be taken by the customer. Sales Representatives were to contact each account and complete an acknowledgement form indicating the Physicians had been notified. For questions regarding this recall customers were to contact their local representative.
Quantity in commerce
1,080 units
Distribution
Worldwide Distribution - USA, Germany, Netherlands, Slovakia, and Spain
Product codes and lots
All Lots, no serial numbers.