Z-2171-2010

IMx Sirolimus Reagent Pack; 100 test pack containing 1 bottle (9.7 mL) Anti-Sirolimus Antibody Coated...

Abbott Medical / initiated 2010-04-30 / Terminated / root cause: manufacturing process

Abbott Laboratories began a recall of IMx Sirolimus Reagent Pack; 100 test pack containing 1 bottle (9.7 mL) Anti-Sirolimus Antibody Coated Microparticles, 1 bottle (9.7 mL) Sirolimus Alkaline... on . The reason FDA records is "IMx Sirolimus reagent lot 802873106 has exhibited an increased frequency of calibration errors". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2171-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
IMx Sirolimus Reagent Pack; 100 test pack containing 1 bottle (9.7 mL) Anti-Sirolimus Antibody Coated Microparticles, 1 bottle (9.7 mL) Sirolimus Alkaline Phosphatase Conjugate, and 1 bottle (10 mL) 4-Methylumbelliferyl Phosphate, 1.2 mM; Produced by Axis-Shield Diagnostics, Ltd., Dundee, UK for Abbott Diagnostics Division, Distributed by Abbott Laboratories, Abbott Park, IL 60064 USA; list 5C91-21 The IMx Sirolimus assay is an in vitro reagent system for the quantitative determination of sirolimus in human whole blood as an aid in the management of renal transplant patients receiving sirolimus therapy.
Product code
NRP Sirolimus Test System
Reason for recall
IMx Sirolimus reagent lot 802873106 has exhibited an increased frequency of calibration errors.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Abbott sent a "Product Recall" letter dated April 30, 2010, to all customers receiving the affected lot of reagent. The accounts were informed of the increased frequency of calibration errors, including: 157 Check 1 Out of Range, 161 Check 5 Out of Range, and 173 Mode 1 deviation Factor High. The accounts were requested to determine if they are currently using or have inventory of lot 802873106, discontinue use of and destroy any remaining inventory of the lot, and retain a copy of the letter for their laboratory records. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688. Customers outside of the U.S., please contact your local Customer Service Representative.
Quantity in commerce
466 kits
Distribution
Worldwide distribution: USA and countries including Puerto Rico, Argentina, Austria, Bahrain, Brazil, Canada, Chile, China, Croatia, Cyprus, Egypt, France, Germany, Greece, Hungary, India, Israel, Italy, Japan, Jordan, Lebanon, Lithuania, Netherlands, Norway, Philippines, Poland, Russia, Panama, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE and Venezuela.
Product codes and lots
Lot number 802873106, Expiry Date May13, 2010