Z-0588-2011

Dimension LIP Flex Reagent Cartridge, Catalog# DF55A

Siemens Healthcare Diagnostics Inc. / initiated 2010-03-10 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension LIP Flex Reagent Cartridge, Catalog# DF55A on . The reason FDA records is "Open Well instability: After calibration, QC shifts high (out of range) after the Flex Well is hydrated. Shifts of up to 20% have been observed in both QC and patient samplesas soon as two hours post hydration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0588-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension LIP Flex Reagent Cartridge, Catalog# DF55A
Product code
CHI Lipase-Esterase, Enzymatic, Photometric, Lipase
Reason for recall
Open Well instability: After calibration, QC shifts high (out of range) after the Flex Well is hydrated. Shifts of up to 20% have been observed in both QC and patient samplesas soon as two hours post hydration.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Recalling firm issued "Urgent Field Safety Notice", dated March 2010, informing users of the problem and to transition to the revised Lipasae method.
Quantity in commerce
21171
Distribution
Nationwide Distribution throughout the US.
Product codes and lots
Lot Number: DE0320, FC0306, GB0342, EA0334, FB0278, DA0293, GB0286, FA0355, EA0362, EB0362.