Dimension LIP Flex Reagent Cartridge, Catalog# DF55A
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension LIP Flex Reagent Cartridge, Catalog# DF55A on . The reason FDA records is "Open Well instability: After calibration, QC shifts high (out of range) after the Flex Well is hydrated. Shifts of up to 20% have been observed in both QC and patient samplesas soon as two hours post hydration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0588-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension LIP Flex Reagent Cartridge, Catalog# DF55A
- Product code
- CHI Lipase-Esterase, Enzymatic, Photometric, Lipase
- Reason for recall
- Open Well instability: After calibration, QC shifts high (out of range) after the Flex Well is hydrated. Shifts of up to 20% have been observed in both QC and patient samplesas soon as two hours post hydration.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Recalling firm issued "Urgent Field Safety Notice", dated March 2010, informing users of the problem and to transition to the revised Lipasae method.
- Quantity in commerce
- 21171
- Distribution
- Nationwide Distribution throughout the US.
- Product codes and lots
- Lot Number: DE0320, FC0306, GB0342, EA0334, FB0278, DA0293, GB0286, FA0355, EA0362, EB0362.