Z-2023-2011

Siemens brand 550 TxT Patient Treatment Table, part number 7346534; Product is manufactured and distributed...

Siemens Medical Solutions USA, Inc. / initiated 2010-03-31 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Siemens brand 550 TxT Patient Treatment Table, part number 7346534; Product is manufactured and distributed by Siemens AG, Medical Solutions, Roentgenstrasse... on . The reason FDA records is "Potential for causing overheating of the power supply during treatment, and potential for unexpected table descent". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2023-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens brand 550 TxT Patient Treatment Table, part number 7346534; Product is manufactured and distributed by Siemens AG, Medical Solutions, Roentgenstrasse 19-21, Kemnath, Germany 95478. The 550 TxT Treatment Table is to be used as an integral part of the Siemens linear accelerator systems to position the patient accurately according to the treatment plan for various radiation therapy treatment techniques.
Product code
JAI Couch, Radiation Therapy, Powered
Reason for recall
Potential for causing overheating of the power supply during treatment, and potential for unexpected table descent.
Root cause tag
design
FDA root cause
Component design/selection
Action
Siemens issued a Customer Information letter dated March 21, 2010. There were two safety issues. There were two occurrences of smoke caused by an overheating of the power supply during patient treatment. Also, there had been a report of an unexpected lowering of the 550TxT table in patient rescue mode. With the released kit that accompanies this letter, Siemens has implemented a minor modification on the 550TxT to include an additional fuse within the power supply circuit to address the first issue described above. They also included an additional modification to address the lowering of the table by the integration of additional relays. As soon as the "move enable" button is released, the table brake will be enabled which is controlled by the implemented relays. No further action is required for the continued operations of the 550TxT table. Further questions regarding this recall please call (925) 246-8407
Quantity in commerce
326 units
Distribution
Worldwide Distribution -- Including USA, states of NJ, WI, NY, and MA.
Product codes and lots
All units of Listing Number: R864406