Z-1974-2010

Horizon Cardiology Hemo Component Schiller Argus Pro PB-1000 Monitor unit. Argus Pro PB 1000 firmware...

Change Healthcare Technologies LLC / initiated 2010-04-15 / Terminated / root cause: software

McKesson Provider Technologies - Medical Imaging Group began a recall of Horizon Cardiology Hemo Component Schiller Argus Pro PB-1000 Monitor unit. Argus Pro PB 1000 firmware versions 2.41+M NIBP and 2.42.01+M. A programmable... on . The reason FDA records is "McKesson has recently discovered that certain Horizon Cardiology Hemo systems were running specific Schiller Argus-Pro PB 1000 firmware versions that might cause a delay in the display of physiological parameters on the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1974-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Change Healthcare Technologies LLC
Product
Horizon Cardiology Hemo Component Schiller Argus Pro PB-1000 Monitor unit. Argus Pro PB 1000 firmware versions 2.41+M NIBP and 2.42.01+M. A programmable diagnostic computer intended for acquiring, digitizing, storing, displaying and reviewing hemodynamic data for use in hospital cardiac catheterization laboratories. The system is cleared for us with a component physiological data collection unit - the Schiller Argus Pro PB 1000 Monitor unit.
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
McKesson has recently discovered that certain Horizon Cardiology Hemo systems were running specific Schiller Argus-Pro PB 1000 firmware versions that might cause a delay in the display of physiological parameters on the Horizon Cariology Hemo Real Time Monitor Screen.
Root cause tag
software
FDA root cause
Software change control
Action
The firm, McKesson, sent an "Advisory Notice" letter dated April 15, 2010, to all customers. The letter describes the problem, products affected, and actions to be taken by customers. The customers were instructed not to use the Emergency Replacement Hemo kit until they receive a replacement Schiller Argus Pro PB-1000 and they have contacted McKesson Support department at 1-866-777-0202. If you have any questions regarding this notice, please call our Customer Support department at 1-0866-777-0202 and ask to speak to your support Manager.
Quantity in commerce
26 units; 19 in operational use, 8 as emergency replacement units.
Distribution
Nationwide distribution: CA, CO, MD, NH, NT, NY, TN, and TX .
Product codes and lots
Argus Pro PB 1000 firmware versions 2.41+M NIBP and 2.42.01+M Serial numbers prefaced with PB-1000: 765.001471, 760.001566, 760.001793, 760.001882, 760.001920, 760.001934, 760.001970, 760.002043, 760.002113, 765.001168, 765.001307, 765.001520, 765.001553, 765.001599, 765.001609, 765.001510, 765.001632, 765.00165, 765.003014, 760.001320, 760,001552, 760.001555, 760,001920, 760.002107, 765.001224, 765.001306, 765.001608.