HeartStart MRx Defibrillator/Monitor with Q-CPR Meter option.The HeartStart MRx monitor/defibrillator has an...

Philips North America LLC / initiated 2010-05-06 / Terminated / root cause: design

Philips Healthcare Inc. began a recall of HeartStart MRx Defibrillator/Monitor with Q-CPR Meter option.The HeartStart MRx monitor/defibrillator has an option (Q-CPR Meter) to provide visual and audible... on . The reason FDA records is "Reduced sensitivity in the force measurement sensor can result in inaccurate feedback to the user on how well he or she is releasing pressure after each chest compression. This can result in incomplete release of chest...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2293-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
HeartStart MRx Defibrillator/Monitor with Q-CPR Meter option.The HeartStart MRx monitor/defibrillator has an option (Q-CPR Meter) to provide visual and audible feedback to the rescuer on the quality of CPR. The Q-CPR option offers real time measurement and corrective feedback on the rate, depth and complete release of compressions, ventilations rate and lack of CPR activity in accordance with current CPR guidelines. The Q-CPR option is contraindicated as follows: *The Q-CPR option is contraindicated for use on neonatal and pediatric patients (under 8 years of age or weighing less than 25 kg). *The Q-CPR option is not for use when CPR is contraindicated.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Reduced sensitivity in the force measurement sensor can result in inaccurate feedback to the user on how well he or she is releasing pressure after each chest compression. This can result in incomplete release of chest compression, which will reduce CPR effectiveness.
Root cause tag
design
FDA root cause
Device Design
Action
Philips issued an "Urgent Medical Device Recall notification/Field Safety Notice" dated April 2010 to all consignees. The letter described the issue, how to identify affected devices, and required actions. The consignees were instructed to remove affected devices from service. Philips will contact consignees to arrange on-site replacement of affected meters with meters containing revised labeling and calibration software. The new meters will indicate that meter recalibration is required after approximately 50,000 compression cycles. Consignees returning devices to Philips for routine recalibration will receive free replacement. All CPR meters will be replaced with re-designed meters when they become available. For questions or assistance in notifying accounts about this recall, contact Philips at 1-800-722-9377.
Quantity in commerce
72 units
Distribution
Worldwide distribution in the countries of USA, Canada, Australia, China, Germany, Iceland, India, Libyan Arab Jamahiri, Netherlands, Saudi Arabia, Spain, Switzerland, United Arab Emirates, and United Kingdom.
Product codes and lots
M3535A/M3536A HeartStart MRx 453564145481 MRx Q-CPR meter 861444, MRx Q-CPR Meter Upgrade 989803162401, MRx Replacement Q-CPR Meter. Lot numbers on distribution list.