The Dimension NTP and LNTP Flex reagent cartridge (N- Terminal Pro- Brain Natriuretic Peptide and N-Terminal...
Siemens Healthcare Diagnostics, Inc. began a recall of The Dimension NTP and LNTP Flex reagent cartridge (N- Terminal Pro- Brain Natriuretic Peptide and N-Terminal Pro brain Natriuretic Peptide (low volume) on . The reason FDA records is "Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0461-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- The Dimension NTP and LNTP Flex reagent cartridge (N- Terminal Pro- Brain Natriuretic Peptide and N-Terminal Pro brain Natriuretic Peptide (low volume)
- Product code
- NBC Test, Natriuretic Peptide
- Reason for recall
- Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Siemens issued an "Urgent Field Safety Notice" dated April 2010, informing users of the problem and instructing them to inspect cartridges for leaking fluid. Cartridges with leaking fluid should be reported to the Siemens Technical Solutions Center. The Technical Solutions Center can be contacted at 800 441 9250.
- Quantity in commerce
- 3364
- Distribution
- Worldwide Distribution: USA, Europe, Australia, China, Japan, Korea, and Saudi Arabia.
- Product codes and lots
- Lot# FD0110 Exp# 04/20/2010, Lot#EB0122, EXP# 05/02/2010, Lot# FC0190, EXP# 07/09/2010, Lot# FC0219, EXP# 08/07/2010, Lot# FB0256, EXP#09/13/2010, Lot# FC0100, EXP#04/10/2010, Lot#EA0122, EXP#02/05/2010, Lot# FB0190, EXP# 07/09/2010, Lot# FB0219, EXP# 08/07/2010, lot# FA0256, EXP# 09/13/2010.