Z-1704-2011

Stryker Stiletto Electrosurgical Probe (E-Probe); Model number: 0270-070-550, 0270-070-551, 0270-070-552...

Stryker Corp. / initiated 2010-02-25 / Terminated / root cause: design

Stryker Endoscopy began a recall of Stryker Stiletto Electrosurgical Probe (E-Probe); Model number: 0270-070-550, 0270-070-551, 0270-070-552, 0270-070-553, 0270-070-554, 0270-070-555... on . The reason FDA records is "The heat shrink, PTFE E-Probe Tip were out of specification (OOS) in the longitudinal change (+5% -20%)". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1704-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker Stiletto Electrosurgical Probe (E-Probe); Model number: 0270-070-550, 0270-070-551, 0270-070-552, 0270-070-553, 0270-070-554, 0270-070-555; Manufactured by Stryker Endoscopy Puerto Rico.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
The heat shrink, PTFE E-Probe Tip were out of specification (OOS) in the longitudinal change (+5% -20%).
Root cause tag
design
FDA root cause
Device Design
Action
All Sales reps were notified of the recall on February 25, 2010. Urgent Device Removal advisory notices were also sent to all consignees.Device Removal letters identified the specific part numbers and lot numbers involved. Letters also list the potential risk to patients as a result of the issue. Customers are asked to inspect their inventory for part and lot numbers and quarantine any affected product. If recalled product is found, customers are to contact Stryker Customer Service at 800-624-4422 and request an RMS to send back recalled units. New units will be provided at no charge. The acknowledgement form is to be completed and returned. Questions or concerns should be directed to 408-754-2000 or e-mail at [email protected].
Quantity in commerce
31, 014 units
Distribution
Worldwide Distribution -- US, Australia, Canada, Germany, Greece, Iberia, Korea, Switzerland, and Newbury.
Product codes and lots
09158FE2, 09161FE2, 09164FE2, 09170FE2, 09174FE2, 09176FE2, 09219FE2, 09226FE2, 09233FE2, 09339FE2, 09264FE2, 09204FE2, 09205FE2, 09210FE2, 09211FE2, 09217FE2, 09120FE2, 09170FE2, 09181FE2, 09181FE2P, 09175FE2, 09209FE2, and 09167FE2.